Chapter 4 Remote Operation and Programming
Command Types
54
Operating and Programming Manual
If the ―:‖ in front of DIAG is omitted, the instrument assumes that you‘ve sent
SYST:DIAG:CONT:RESET and will generate an error.
*RST Reset Response
The 5071A reset response (Initialization state) is the configuration acquired at
program reset. Table 4-2 summarizes this state. Once a configuration has been
saved, it will persist even if power is removed or until it is reprogrammed (via
SCPI commands or through the front-panel menus).
Table 4-2. *RST Reset Conditions
Subsystem
Command/Parameter
State
DIAGnostic
:LOG:VERBosity
DISable
DISPlay
:ENABle
ON
[SOURce]:PTIMe
:SYNC
OFF
[SOURce]:ROSCillator
:STEer
0.0
SYSTem
:REMote
ON
The IEEE 488.2 *RST command returns the instrument to a specified state
optimized for remote operation. Use *CLS to clear the status registers and the
SCPI error queue.
Summary of Contents for 5071A
Page 2: ......
Page 28: ...Preface xxvi Operating and Programming Manual ...
Page 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Page 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Page 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Page 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Page 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Page 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Page 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Page 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Page 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Page 145: ...____________________________________________ Chapter 7 Installation Installation ...
Page 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Page 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Page 187: ...____________________________________________ Appendix B Glossary Glossary ...
Page 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...