Chapter 5 Command Reference
Description of SCPI Commands
86
Operating and Programming Manual
PTIMe:LEAPsecond[:STATe] OFF|0|ON|1
This command activates or cancels a programmed leap second.
When STATe is turned ON, a leap second is scheduled for the last specified
values of DURation and MJDate.
The function of this command is not affected by *RST.
PTIMe:LEAPsecond[:STATe]?
This query indicates whether a leap second is pending.
Response format = <0 or 1>
Returns 1 if pending, 0 if none pending.
PTIMe:SLEW <numeric_value>
This command adjusts the epoch of the clock with the supplied parameter.
The 5071A must be in REMote mode to execute this command.
One parameter is required, accepts ―S‖ suffix.
Parameter range = -0.5 S to 0.5 S., precision = 50 ns.
This is an event.
The function of this command is not affected by *RST.
PTIMe:SLEW? MIN|MAX
This query returns the smallest and largest slew value which can be used.
Response format = <NR3>.
PTIMe:STANdby OFF|0|ON|1
This command enables or disables STANDBY mode.
The 5071A must be in REMote mode to execute this command.
Disabling standby initiates normal operation, if possible. (This command is
equivalent to the 5061B ―cesium off‖ switch.)
The function of this command is not affected by *RST.
Summary of Contents for 5071A
Page 2: ......
Page 28: ...Preface xxvi Operating and Programming Manual ...
Page 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Page 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Page 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Page 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Page 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Page 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Page 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Page 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Page 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Page 145: ...____________________________________________ Chapter 7 Installation Installation ...
Page 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Page 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Page 187: ...____________________________________________ Appendix B Glossary Glossary ...
Page 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...