Chapter 4 Remote Operation and Programming
Command Types
Operating and Programming Manual 51
Abbreviated Commands
The command syntax shows most commands as a mixture of upper and lower
case letters. Upper case letters indicate the abbreviated spelling for the command.
For shorter program lines, send the abbreviated form. For better program
readability, you may send the entire command. The 5071A accepts either
command form and is not case sensitive.
For example, if the command syntax shows DIAGnostic, then DIAG and
DIAGNOSTIC are both acceptable forms. Other forms of DIAGnostic, such as
DIAGN or DIAGNOST will generate an error. You may use upper or lower case
letters. Therefore, DIAGNOSTIC, diagnostic, and DiAgNoStIc are all acceptable.
Implied Channel
Some commands contain channel numbers in square brackets: [1|2]. The
brackets indicate that the same programming sequence can be used for both
channels, one of which must be specified. If a channel number is not specified, it
will default to 1.
Implied Commands
Implied commands appear in square brackets ([ ]) in the command syntax. (The
brackets are not part of the command and are not sent to the instrument.)
Suppose you send a second level command but do not send the preceding implied
command. In this case, the instrument assumes you intend to use the implied
command and responds as if you had sent it.
Examine the portion of the DIAGnostic subsystem shown below:
SOURce]
:PTIMe
[:TIME] <hour>,<minute>,<second>
[:TIME]? [MIN|MAX],[MIN|MAX],[MIN|MAX]
The root command SOURce is an implied command (as is :TIME). To set the
instruments time to 9:30 am, you can send either of the following command
statements:
SOURce:PTIMe:TIME 09,30,00
or
PTIMe 09,30,00
Summary of Contents for 5071A
Page 2: ......
Page 28: ...Preface xxvi Operating and Programming Manual ...
Page 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Page 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Page 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Page 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Page 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Page 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Page 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Page 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Page 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Page 145: ...____________________________________________ Chapter 7 Installation Installation ...
Page 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Page 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Page 187: ...____________________________________________ Appendix B Glossary Glossary ...
Page 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...