Safety Considerations and Warranty
Operating and Programming Manual
iii
Indicates terminal is connected to chassis when
such connection is not apparent.
Indicates Alternating current.
Indicates Direct current.
Rack Mounting
The Model 5071A Unit is designed for mounting in a standard 19-inch (48.26 cm) rack.
Follow the rack manufacturer's instructions for mounting the Model 5071A Unit while
observing the following guidelines:
Elevated Operating Temperature: If the Model 5071A Unit is installed in a
closed or multi-unit rack assembly, the ambient temperature of the rack
environment may be greater than the Model 5071A Unit's Maximum Operating
Temperature of 50°C/122°F. Install the Model 5071A Unit in an environment
that is compatible with the Model 5071A Unit 's operating temperature range,
which is 0 °C to 50 °C, or 32 °F to 122 °F
Mechanical Loading: Mount the Model 5071A Unit so as to avoid uneven
mechanical loading that could cause hazardous conditions.
Circuit Overloading: Observe the power ratings on the Model 5071A Unit 's
nameplate and the additional load the Model 5071A Unit may place on the
supply circuit. Overloading the supply circuit may adversely affect the supply
wiring and over-current protection.
Reliable Earthing: Maintain reliable earthing (grounding) of rack mounted
equipment. Pay particular attention to supply connections other than direct
connections to the branch circuit (e.g., use of power strips).
CAUTION
Damage to equipment, or incorrect measurement data, may result from failure to heed a
caution. Do not proceed beyond a
CAUTION
sign until the indicated conditions are fully
understood and met.
Whenever it is likely that the protection has been impaired, the instrument must be made
inoperative and be secured against any unintended operation.
If this instrument is to be energized via an autotransformer (for voltage reduction) make
sure the common terminal is connected to the earthed pole terminal (neutral) of the power
source.
Instructions for adjustments while covers are removed and for servicing are for use by
service-trained personnel only. To avoid dangerous electric shock, do not perform such
adjustments or servicing unless qualified to do so.
----------------------------------------------------------------------------------------------------
Or
Summary of Contents for 5071A
Page 2: ......
Page 28: ...Preface xxvi Operating and Programming Manual ...
Page 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Page 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Page 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Page 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Page 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Page 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Page 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Page 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Page 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Page 145: ...____________________________________________ Chapter 7 Installation Installation ...
Page 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Page 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Page 187: ...____________________________________________ Appendix B Glossary Glossary ...
Page 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...