6.11. Cleaning the solenoid valve
Disassemble
the solenoid
valve
The solenoid valves are mounted at the bottom of the flow cell
block. The solenoid valve should be disassembled if it does not
switch anymore or if it is clogged.
1
Switch off the instrument according to instructions in
Operation for Maintenance, p. 45
A
B
Valve 1 switches the sample flow
through the photometer on and off.
Valve 2 is used for adding the
Reagent Oxycon on-line KI.
2
Loosen the nut (A).
3
Remove the solenoid coil (B) from
the valve body (C).
B
A
A
B
C