46
A-96.250.581 / 220616
AMI Codes-II CC
Maintenance
6.3.
Refill or replace Reagents
The liquid level in the containers is monitored.
The following messages are displayed:
NOTICE:
Before refilling the reagents, rinse the container with
demineralized water.
Reagent
consumption
The 2 liter reagent container for DPD & Buffer will last for 19 days of
operation (with measurement interval of 4 minutes). The 2 liter re-
agent container for KI will last for 66 days of operation (with mea-
surement interval of 20 minutes). The provided reagent set (for 8
containers) therefore lasts for 5 respectively 17 months of opera-
tion.
As the reagent consumption is not linear find further examples be-
low:
WARNING
Health hazard
Some reagents can cause severe burns or eye damage.
For safe handling of the reagents you must read and under-
stand the Material Safety Data Sheets (MSDS), see
Container almost
empty
Maintenance E065 - Reagents low and the
remaining reagent volume in %
(starting at 17 % = 340 ml).
Container empty
Error E022 - Reagent empty
Measuring interval
Duration of DPD/Buffer
Duration of KI
fc
tc
one canister
filling
reagent set one canister
filling
reagent set
2 minutes
10 minutes
~ 14 days
3 months
~ 33 days
8 months
4 minutes
20 minutes
~ 19 days
5 months
~ 66 days
17 months