AMI Codes-II CC
Program List and Explanations
A-96.250.581 / 220616
95
5.3.1.2
Case Temp. high
: Set the alarm high value for temperature of elec-
tronics housing. If the value rises above the programmed value
E013 is issued.
Range: 30–75 °C
5.3.1.3
Case Temp. low
: Set the alarm low value for temperature of elec-
tronics housing. If the value falls below the programmed value
E014 is issued.
Range: -10 to +20 °C
5.3.1.4
Sample Flow:
Define at which sample flow a flow alarm should be
issued.
5.3.1.4.1
Flow Alarm:
Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will al-
ways be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
NOTICE:
Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.4.2
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 100–600 B/s
5.3.1.4.36
Alarm Low:
If the measuring values falls below the programmed
value E010 will be issued.
Range: 5–80 B/s
5.3.1.5
Temperature:
Only available if pH option is installed. Define the
measuring value, which should issue an alarm high respectively
low.
5.3.1.5.1
Alarm High:
If the sample temperature rises above the pro-
grammed value E007 is issued.
Range: 30–70 °C
5.3.1.5.26
Alarm Low:
If the sample temperature falls below the programmed
value E008 is issued.
Range: 0–20 °C