Maintenance
A-96.250.581 / 220616
59
6.9.
Maintenance of pH sensor
Clean
pH sensor
1
Remove the pH sensor [B] from the flow cell.
2
Unscrew and remove the connector [A] from the pH sensor.
Prevent the connectors from getting wet
3
If necessary wipe the pH sensor shaft and the green tip cau-
tiously with a soft, clean, and damp paper tissue.
4
Remove grease with a tissue moistened with alcohol.
CAUTION: Alcohol is inflammable.
5
If the sensor is very dirty put it into 1% diluted hydrochloric acid
for 1 min.
CAUTION! hydrochloric acid is corrosive!
6
Rinse the pH sensor with clean water.