32
A-96.250.581 / 220616
AMI Codes-II CC
Installation
3.8.2
Relay 1 and 2
NOTICE:
Max. load 1 A/250 VAC
Relay 1 and 2 can be configured as normally open or as normally
closed. Standard for both relays is normally open. To configure a
Relay as normally closed, set the jumper in the upper position.
NOTICE:
Some error codes and the instrument status may
influence the status of the relays described below.
For programming see menu Installation
.
Relay
config.
Terminals
Jumper
pos.
Description
Relay configuration
Normally
Open
6/7: Relay 1
8/9: Relay 2
Inactive (opened) during
normal operation and
loss of power.
Active (closed) when a
programmed function is
executed.
Normally
Closed
6/7: Relay 1
8/9: Relay 2
Inactive (closed) during
normal operation and
loss of power.
Active (opened) when a
programmed function is
executed.
6
0V
7
6
0V
7
A
B
Jumper set as normally open (standard setting)
Jumper set as normally closed