AMI Codes-II CC
Maintenance
A-96.250.581 / 220616
61
Dismount
pump tubes
The pump tube can easily be dismounted and mounted.
Proceed as follows:
1
Switch off the instrument according to instructions in
Operation for Maintenance, p. 45
2
Open the occlusion frame (B) by turning it counter-clockwise.
3
Remove the occlusion frame (B) from the rotor (C) by pulling
the complete occlusion frame out of the holder.
4
Dismount all reagent tubes from the occlusion frame.
5
Exchange the occlusion frame.
6
Install all reagent tubes.
7
Lock the occlusion frames.
8
Start fill system.
A
B
C
D
E
F
Pump housing
Occlusion frame open
Rotor
Pump tube
Pump inlet
Pump outlet
A
B
D
C
F
E