AMI Codes-II CC
Maintenance
A-96.250.581 / 220616
49
4
Close the canister with the screw cover and tighten it well.
5
Mix the demineralized water and the reagents well.
6
Put the canister [E] into the holder [H].
7
Remove the screw cover and insert the suction lance [C] com-
bined with level detector [B] and tighten the screw cover [A].
Prepare
Oxycon
On-line Buffer
1
Rinse the canister [F] labelled “OXYCON ON LINE Buffer” solu-
tion with demineralized water.
2
Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3
Slowly pour the content of one bag of buffer Oxycon On-line
Buffer into the canister.
Avoid splashing!
4
Close the canister with the screw cover and tighten it well.
5
Mix the demineralized water and the reagents well.
6
Put canister [F] into holder [H].
7
Remove the screw cover and insert the suction lance [C] com-
bined with level detector [B] and tighten the screw cover [A].
Prepare
Oxycon
On-line KI
1
Rinse the canister [G] labelled “OXYCON ON LINE KI” solution
with demineralized water.
2
Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3
Slowly pour the content of one bag of Oxycon On-line KI into
the canister.
Avoid splashing!
4
Close the canister with the screw cover and tighten it well.
5
Mix the demineralized water and the reagents well.
6
Put canister [G] into holder [H].
7
Remove the screw cover and insert the suction lance [C] com-
bined with level detector [B] and tighten the screw cover [A].
8
Fill reagent system. See