48
A-96.250.581 / 220616
AMI Codes-II CC
Maintenance
Canister set up
6.3.1
Reagents for measuring Free Chlorine and Total Chlorine
To prepare 2 L reagents the following chemicals are necessary:
1 x 50 ml of Oxycon On-line DPD
1 x 240 gr of Oxycon On-line Buffer
1 x 60 gr of Reagent Oxycon On-line KI
6 liters demineralized water
NOTICE:
To mix the reagents use a screw cover without the
integrated suction lance and level detector.
Prepare
Oxycon
On-line DPD
1
Rinse the canister [E] labelled “OXYCON ON LINE DPD-Re-
agent” with demineralized water.
2
Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3
Slowly pour the content of one bottle of concentrate Oxycon
On-line DPD (50 ml) into the canister.
Avoid splashing!
A
B
C
D
E
F
G
H
Screw cover
Level detector
Suction lance
2 L mark
Canister Oxycon
on-line DPD
Canister Oxycon
on-line Buffer
Canister Oxycon
on-line KI
Holder
OXYCON ON-LINE
Pufferlösung
Tampon
Buffer Solution
2L
OXYCON ON-LINE
DPD-Reagenz
DPD Réactif
DPD Reagent
2L
2L
OXYCON ON-LINE
KI-Lösung
KI-Solution
A
B
C
E
F
G
D
H