58
A-96.250.581 / 220616
AMI Codes-II CC
Maintenance
6.8.2
Assemble the Flow Cell
1
Replace the gasket [D] before reassembling the flow cell.
NOTICE:
A film of teflon paste (e.g. Fomblin from Solvay
Solexis) on the gaskets improves tightness and life time.
2
Push the overflow tube [B] through the flow cell block as far as it
reaches the drain.
3
Install the constant head tube [C] onto the flow cell block.
4
Put the cover onto the constant head.
5
Align the overflow tube with the upper level mark.
A
B
C
D
E
Constant head cover
Overflow tube
Constant head tube
Gasket
Flow cell block
Level
A
B
D
C
E