4-42
17 364 02 0320
Operation
4.6.2
Start-up
• Check that there are no bubbles in the handpiece.
If bubbles are visible under the coupling diaphragm, proceed as follows:
Position the handpiece in the handpiece holder. This ensures that air bubbles will
always be sucked out of the handpiece.
• Leave the handpiece in this position for approx. 3 minutes until the suction
procedure has finished.
• To work in F-SW mode, set the shock energy to an initial value of 0.1 mJ/mm
2
.
The maximum energy level corresponds to an energy flux density of 0.55 mJ/mm
2
.
• Optional: To work in C-ACTOR II mode, set the pulse energy to an initial value
of 0.03 mJ/mm
2
.
The maximum energy level corresponds to an energy flux density of 1.24 mJ/mm
2
.
NOTE
The highest permitted frequency is always set when an energy level is selected
(see
t
able
4-10
and
t
able
4-11
). This frequency can be reduced manually.
• Activate the trigger button.
The trigger button
functions as an on/off switch when it is pressed briefly (< 1.5 s).
Pressing it for longer (> 1.5 s) causes it to function as a tip switch, i.e. the shocks will
continue until the button is released.
4.6.3
Functional checks
Perform the following functional checks after the system has been installed:
• Check the control unit and handpieces for damage.
•
Start up the DUOLITH
®
SD1 (see
C
hapter
4.6.2 s
tart
-
up
)
.
• Set the energy level in F-SW mode to 0.2 mJ/mm
2
.
• Optional: Set the energy level in C-ACTOR II mode to 0.69 mJ/mm
2
.
• Reset the actual number of shocks on the parameter display of the control panel
with the Reset button
.
•
Release shocks with a shock frequency of 4 Hz.
• Check that the triggered shocks are correctly counted on the treatment shock
counter.
NOTE
If necessary, the functional capability of the F-SW handpiece can be checked with
the aid of special pressure sensors (see
C
hapter
7.1 a
CCessOries
).
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...