![Storz DUOLITH SD1 R-SW Operating Manual Download Page 104](http://html1.mh-extra.com/html/storz/duolith-sd1-r-sw/duolith-sd1-r-sw_operating-manual_1383000104.webp)
5-4
17 365 02 0918
Status messages and tr
ouble-shooting
Printer not ready
Connect and switch on the printer.
USB stick was not recognised
Remove the USB stick, then switch off and
restart the device. Reinsert the USB stick.
Check that there is software on the USB stick.
If the fault persists, check that the USB stick
supports the USB V1.1 protocol. If it does
not, replace the USB stick.
5.2
Trouble-shooting
CAUTION!
Unplug the mains cable from the device before you carry out
any maintenance work!
Fault description
Possible cause
Corrective action
System does not work
Power failure
Defective mains fuse
Defective mains plug
Check the power supply.
Replace the fuses.
Replace the mains cable.
No compressed air
supply
Defective foot switch
connecting cable
Leaks in handpiece
cable or cable not
properly connected
Pressure setting
incorrect
Clogged compressor
air filter
Check the connecting cable
and change the foot switch if
necessary.
Check the connecting cable and
change the handpiece if necessary.
Check the pressure setting.
Check the compressor air filter
and replace it, if necessary.
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...