![Storz DUOLITH SD1 R-SW Operating Manual Download Page 103](http://html1.mh-extra.com/html/storz/duolith-sd1-r-sw/duolith-sd1-r-sw_operating-manual_1383000103.webp)
5-3
17 365 02 0918
Status messages and tr
ouble-shooting
F-SW: pump temperature too high
Acknowledge message,
further treatment is possible once the pump
temperature has returned to permitted
values.
F-SW: therapy head overtemperature
Acknowledge message,
further treatment is possible once the
therapy head temperature has returned to
permitted values.
F-SW: water temperature sensor
failure
Restart the system.
Inform your Service centre if the fault
continues.
Shock wave limit for current
handpiece reached
Shock wave limit for current handpiece
reached. Replace the handpiece.
F-SW: CTU fault: fault switch
Charging unit not ready
Acknowledge message.
Call your Service centre if the fault continues
after a reset.
F-SW: CTU fault: OK missing
Charging unit not ready
Acknowledge message.
Call your Service centre if the fault continues
after a reset.
F-SW: CTU fault: divider disconnected
Charging unit not ready
Acknowledge message.
Call your Service centre if the fault continues
after a reset.
F-SW: CTU fault: hv low
Charging unit not ready
Acknowledge message.
Call your Service centre if the fault continues
after a reset.
F-SW: CTU fault: timeout
Charging unit not ready
Acknowledge message.
Call your Service centre if the fault continues
after a reset.
F-SW: CTU fault: overtemperature
Charging unit not ready
Acknowledge message.
Call your Service centre if the fault continues
after a reset.
F-SW: CTU fault: hv set beyond spec.
Charging unit not ready
Acknowledge message.
Call your Service centre if the fault continues
after a reset.
F-SW: CTU fault: interlock
Charging unit not ready
Acknowledge message.
Call your Service centre if the fault continues
after a reset.
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...