3-9
17 363 02 0320
System description
Changing the mains voltage
•
Move the switch using a pointed object (such as a flat-blade screwdriver) until
the required value is under the arrowhead.
Fig. 3-9 Mains voltage set to 115 V
• Adjust the fuse value of the mains fuse according to the selected mains voltage
(see c
hapter
6.5 f
use
replaCeMent
).
Connecting the electrical power supply
• Connect the mains cable to the mains connection on the rear of the device.
1 Mains connector
1
Fig. 3-10 Mains connector
• Insert the mains cable into the electrical socket.
3.4
Transport
• Make sure there are no objects on the device.
• Remove the handpiece and, if supplied, the ultrasound probe and foot switch.
• Disconnect the mains plug.
• Remove the potential equalisation connector.
ATTENTION!
Incorrect handling during transport will damage the device!
• Only hold the device at the handle when transporting.
Do not use either the handpiece holder or the monitor for pushing or pulling
the device.
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...