3-15
17 363 02 0320
System description
3.6.4
F-SW SEPIA handpiece
Focused shock waves that are concentrated on a focal zone outside the handpiece are
administered over the F-SW handpiece into the body at the treatment zone
.
1
6
4
1 Display
2 Trigger button
3 Clamping ring
4 Fixing screws
5 Coupling diaphragm
6 Color ring
2
3
5
Fig. 3-22 F-SW handpiece
The coupling diaphragm is fixed by a clamping ring and 3 fixing screws. It can only be
opened from authorised personnel with special tools.
The penetration depth of the shock wave can be adjusted by using different stand-off
devices.
Fig. 3-23 F-SW handpiece
without stand-off device with stand-off device I with stand-off device II
Therapeutically effective
penetration depth 5 MPa
0 - 125 mm
0 - 105 mm
0 - 90 mm
Depth of
focal zone
35 - 65 mm
15 - 45 mm
0 - 30 mm
15 mm
30 mm
50 mm
Fig. 3-24 Depth of therapeutical effect
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...