6-12
17 366 02 0320
Cleaning, maintenance, overhaul
6.9
Service life
Device
The average expected service life (MTTF)
according to IEC 60601-1:2005 + A1:2012 / EN 60601-
1:2006 + A1:2013 is:
DUOLITH SD1
– approx. 15,000 operating hours
F-SW handpiece
– approx. 5 million shocks
Prerequisite:
The coil is a wear part. After 2 million shocks the coil needs
to be exchanged.
R-SW handpiece
– approx. 5 million shocks for the handpiece
– approx. 1 million shocks for the shock transmitters.
Prerequisite:
Every 1 million shocks the R-SW handpiece should be
overhauled (see Chapter
6.4 r-sW
handpieCe
Overhaul
).
V-ACTOR
– approx. 5 million pulses
suction cup
– 100 cleaning and disinfecting cycles
Exceeding the service life can be expected to result in a failure of the devices and the
accessories. This also applies for the handpieces.
No warranty claims shall be accepted beyond the information given in
C
hapter
9.1
W
arranty
fOr
the
duOlith
®
sd1
.
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...