6-4
17 366 02 0320
Cleaning, maintenance, overhaul
6.2.2.1
Draining the water circuit
The water circuit must be drained if the device will not be used for several weeks.
• Make sure that the device is standing on a level surface.
• Press
to load F-SW operating mode.
• Press
.
• In the open menu list, select the
function.
• Connect the water bag to the DUOLITH
®
SD1 and put it on the floor when
prompted to do so.
The water hose connection on the DUOLITH
®
SD1 is located on the rear of
the module.
Fig. 6-1 Water hose connection
•
Confirm by pressing OK.
– A progress display indicating that the water circuit is draining is displayed at the
bottom left of the display.
In order to drain the water circuit of the DUOLITH
®
SD1 completely, the water in the
handpiece must be drained as well.
A prompt to this effect appears on the display:
• Hold the handpiece in an upright position above the device to optimise the water
flow from the device.
• Make sure that the coupling diaphragm of the handpiece is pointing upwards.
Fig. 6-2 Draining handpiece
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...