Status messages and tr
ouble-shooting
Fault description
Possible cause
Corrective action
No R-SW power
output
No F-SW power
output
Blocked or worn
projectile
2 projectiles in the
handpiece
F-SW handpiece
defective
Malfunction in control
device
Dismantle the R-SW handpiece,
clean the guide tube and
projectile, overhaul the handpiece
(see
C
hapter
6.4 r-sW
handpieCe
Overhaul
)
Dismantle the R-SW handpiece
and remove the excess projectile.
Replace the handpiece.
Call your Service centre.
No F-SW power
output
Handpiece has not
been recognised
Check that the blue screw is
screwed to the correct tightness.
Shock triggering noise
changes after several
shocks
Air in handpiece
Hold handpiece vertically with
coupling diaphragm downwards
so that air is sucked out.
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...