6-7
17 366 02 0320
Cleaning, maintenance, overhaul
6.3
Modul R-SW
6.3.1
Cleaning the R-SW handpieces
Regular cleaning ensures perfect hygiene and operation of the R-SW handpiece.
The handpiece, in particular the shock transmitter, must be thoroughly cleaned and
disinfected after each therapy session.
CAUTION!
Flammable and volatile cleaning and disinfecting agents may
generate an explosive atmosphere.
• Disconnect the handpiece from the control unit before starting any
cleaning or maintenance work.
ATTENTION !
It is essential that no fluid be permitted to penetrate either the device or its tubing.
• Clean the coupling gel off the handpiece using a cleaning agent that is suitable for
surface cleaning.
• Disinfect the handpiece with an alcohol-based disinfectant that is suitable for
surface cleaning (follow the manufacturer’s instructions).
For cleaning the handpiece please read the
separate
O
perating
M
anual
Of
the
handpieCe
.
6.3.2
Cleaning the V-ACTOR
The handpiece must be thoroughly cleaned and disinfected after each therapy session.
• Clean the remaining massage oil off the handpiece using a damp cloth.
• Disinfect the handpiece with a surface disinfectant.
• Clean the shock transmitters thoroughly after each use. Unscrew the shock
transmitter screw cap and remove the shock transmitter insert if necessary.
• The shock transmitters can be cleaned using the usual cleaning agents and
disinfectants.
NOTE
We recommend that the shock transmitters be cleaned in an ultrasonic bath.
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...