6-10
17 366 02 0320
Cleaning, maintenance, overhaul
6.4
R-SW handpiece overhaul
Shock waves are generated mechanically. Due to the effects of friction, the handpiece
components are continuously exposed to mechanical stress, which will cause minor
wear.
NOTE
The R-SW handpiece should be overhauled about every 1,000,000 shocks.
This can be done quickly and easily by the user of the device. All that is required
is the overhaul kit, which includes all required wear parts.
For overhauling the R-SW handpiece please read he
separate
O
perating
M
anual
Of
the
r-sW
handpieCe
.
6.5
Fuse replacement
The mains fuse holder is located on the rear of the DUOLITH
®
SD1.
• Push the slide of the mains fuse holder upwards and release the holder from the
housing.
Fig. 6-5 Mains connector, fuse holder
• Pull the old fuses out of the mains fuse holder.
• Replace the fuses
(T 6.3 AH / 250 VAC at 115 VAC mains input voltage respectively
T 4 AH / 250 VAC at 230 VAC mains input voltage).
• Push the mains fuse holder back into the opening until it engages.
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...