8-7
17 368 02 0320
Technical data, conformity and labelling
Guidelines and manufacturer's declaration –
Resistance to emitted electromagnetic interference
The DUOLITH
®
SD1 model is intended for operation in the electromagnetic environment specified below.
The customer or the user of the DUOLITH
®
SD1 should ensure that it is used in such an environment.
Immunity tests
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidelines
Electrostatic
discharge (ESD) acc.
to IEC 61000-4-2
±8 kV contact
discharge
±15 kV air
discharge
±8 kV contact
discharge
±15 kV air
discharge
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient
disturbances /
bursts according to
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surges according to
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage drops, short
interruptions and
voltage variations on
power supply input
lines according to
IEC 61000-4-11
< 5% U
T
(> 95% drop in
U
T
) for ½ period
40% U
T
(60% drop in U
T
)
for 5 periods
70% U
T
(30% drop in U
T
)
for 25 periods
< 5% U
T
(> 95% drop in
U
T
) for 5 s
< 5% U
T
(> 95% drop in
U
T
) for ½ period
40% U
T
(60% drop in U
T
)
for 5 periods
70% U
T
(30% drop in U
T
)
for 25 periods
< 5% U
T
(> 95% drop in
U
T
) for 5 s
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the DUOLITH
®
SD1 requires
continued operation during power mains
interruptions, it is recommended that
the DUOLITH
®
SD1 be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A/m
30 A/m
The mains frequency magnetic fields should
be those of a typical business or hospital
environment.
NOTE U
T
is the mains alternating voltage prior to application of the test level.
Summary of Contents for DUOLITH SD1 R-SW
Page 2: ...II 17 360 02 0320 Preface...
Page 12: ...XII 17 360 02 0320 Preface...
Page 13: ...1 General Safety Information...
Page 19: ...2 Principles...
Page 24: ...2 6 17 362 02 0320 Principles empty page...
Page 25: ...3 System Description...
Page 49: ...4 Operation...
Page 100: ...4 52 17 364 02 0320 Operation...
Page 101: ...5 Status Displays and Trouble shooting...
Page 106: ...5 6 17 365 02 0918 Status messages and trouble shooting empty page...
Page 107: ...6 Cleaning Maintenance Overhaul...
Page 119: ...7 Accessories...
Page 122: ...7 4 17 367 02 0320 Accessories...
Page 123: ...8 Technical data conformity and labelling...
Page 137: ...9 Warranty and service...
Page 139: ...10 Ultrasound optional...