6-10
122998-741
Operator Manual
Routine Maintenance
6.5.3 Clean Water Inlet
Strainer
Strainer, located in the incoming water line to vapor condenser, must
be cleaned once every three months.
1.
Set
POWER-OFF/STANDBY
switch to
STANDBY
. Lock
disconnect switch in OFF position and close building supply
valves.
2.
Pull out strainer screen from body.
3.
Using a wire brush or steel wool, scrape all rust and residue from
strainer screen and body. Ensure all perforations are clear, by
poking open with a wire. Replace screen if damaged or cor-
roded.
4.
Verify no dirt or other particles remain in strainer body. Insert
screen into strainer body.
5.
Using pipe joint sealing compound, replace and tighten hex
cap.
6.
Verify all pipe connections are tight and leak free after reassem-
bly.
7.