5-8
122998-741
Operator Manual
Cycle and Control Value Programming
5.3.4 Rinse - All Units
1.
Press
CHANGE VALUES
touch pad. Rinse 1 phase values
screen appears:
CYCLE 10
SP=180.0 F
RINSE 1
T=01M 30S
__Indicates flashing position.
To enter First Rinse temperature set point and phase time, press
CURSOR arrows
to select position and
VALUE arrows
to select
desired number (0-9). Temperature set point is input as any
number value to 1/10th of a degree. Phase time is input as
minutes and seconds within a range of 0-49 minutes and 0-99
seconds. Temperature is adjustable from 70 to 185ºF (22 to
85ºC).
2.
Press
CHANGE VALUES
touch pad. Rinse 1 Temperature Guar-
antee option screen appears:
GUAR. RINSE 1 TEMP.
NO
__Indicates flashing position.
Press
VALUE arrow
(either up or down) to toggle between YES
and NO answer. Selecting NO starts Rinse 1 phase time count
down at beginning of phase. Selecting YES guarantees that
Rinse 1 phase time will count down only when water temperature
is equal to or greater than programmed set point.
NOTE: With Temperature Guarantee feature enabled, phase
time will not count down unless set point is reached. It is
important that set point be an attainable value. Guaranteed
temperature is shown by a G before set point (GSP instead of
SP).
3.
Press
CHANGE VALUES
touch pad. Rinse 2 phase values
screen appears:
CYCLE 10
SP=180.0 F
RINSE 2
T=MM:SS
__Indicates flashing position.
To enter Rinse 2 temperature set point and phase time, press
CURSOR arrows
to select position and
VALUE arrows
to select
desired number (0-9). Temperature set point is input as any
number value to 1/10th of a degree. Phase time is input as
minutes and seconds within a range of 0-49 minutes and 0-99
seconds. Temperature is adjustable from 70 to 185ºF (22ºC to
85ºC).
4.
Press
CHANGE VALUES
touch pad. Rinse 2 Temperature Guar-
antee option screen appears:
GUAR. RINSE 2 TEMP.?
NO
__Indicates flashing position.