4-14
122998-741
Operator Manual
Operating Instructions
pH TOO HIGH
MIXING WATER
7.
Sump water is pumped to drain. If temperature is guaranteed,
actual sump temperature alternates with guaranteed set temper-
ature:
CYCLE 10
(G=)128.5 F
F. RINSE DRAIN
1:00
4.3.15 Exhaust Phase
1.
Washer stands idle for programmed time allowing residual
vapors to exhaust from wash chamber to room:
CYCLE 10
EXHAUST
1:30
2.
Printer prints time at completion of phase:
EXHAUST
10:38:05A
**************************************
NOTE: Customer must connect exhaust vent to building exhaust
system to prevent directly exhausting vapors to room.
4.3.16 Cycle Complete
1.
Alarm buzzer sounds and an operator instruction is displayed.
Alarm buzzer can be silenced by pressing
ALARM REPLY
touch pad or opening chamber door(s). Instruction remains on
screen until door(s) is opened.
PLEASE OPEN DOOR
AND REMOVE THE LOAD
... then printer records:
COMPLETE
10:39:35A
MAX WASH TEMP= 141.5F
MAX RINSE TEMP= 180.8F
CYCLE TIME= 0:23:15
======================
= READY TO UNLOAD =
=======================
2.
Cautiously open chamber door to vent remaining steam vapors.
Allow chamber to cool a few minutes before removing load. Dis-
play screen returns to cycle menu:
CYCLE 9
CYCLE 10
CYCLE 11
CYCLE 12
__Indicates flashing position.
...and printer records:
DOOR OPENED 10:42:10A
************************************
WARNING – BURN HAZARD:
Wear appropriate Personal
Protective Equipment (PPE)
including gloves and face
protection, open door slowly,
and allow chamber and load
to cool when cycle is com-
plete. Hot steam may escape
through door opening if door
is fully opened after cycle is
complete.
WARNING – SLIPPING HAZ-
ARD: To prevent slips, keep
floor dry. Promptly clean up
any spills or drippage. If
spills or drippage are deter-
gents or other chemicals, fol-
low safety precautions and
handling procedures set forth
on detergent or chemical
label and/or Material Safety
Data Sheet (MSDS).