4-10
122998-741
Operator Manual
Operating Instructions
4.3.9 Rinse 1 Phase
1.
Hot water, from building supply, fills sump. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature.
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time will only count
down when solution temperature is equal to or greater than set
point.
CYCLE 10
(G=)137.4 F
RINSE 1 FILL
2:00
2.
Rinse 1 water is recirculated through oscillating jet system for
programmed amount of time. If temperature is guaranteed,
actual sump temperature alternates with guaranteed set temper-
ature.
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time will only count
down when solution temperature is equal to or greater than set
point.
Display shows:
CYCLE 10
(G=)140.0 F
RINSE 1
5:00
3.
Printer records time and solution temperature in chamber sump
at beginning and end of recirculation:
RINSE 1
10:21:55A
141.5
10:26:55A 140.2
****************************************
4.3.10 Controller-Driven pH
Neutralization (Option)
1.
pH analyzer is initialized:
STANDBY
NEUTRALIZING SUMP
2.
pH analyzer verifies that value is within limits:
CHECKING pH
pH = XX.XXX
3.
If pH is too low, while injecting, acid neutralizer is injected to
compensate:
pH TOO LOW
INJECTING ACID NEUT.
4.
Water-Acid neutralizer solution is mixed:
pH TOO LOW
MIXING WATER
5.
If pH is too high, while injecting, alkaline neutralizer is injected to
compensate:
pH TOO HIGH
INJECTING ALK. NEUT.
WARNING – BURN HAZARD:
Water discharge may be
extremely hot.