5-4
122998-741
Operator Manual
Cycle and Control Value Programming
5.3.1 Pre-Wash - All Units
1.
Press
CHANGE VALUES
touch pad. Pre-Wash screen appears:
CYCLE 10
HTW
PRE-WASH T=MM:SS
__Indicates flashing position.
2.
To enter Pre-Wash phase time, press
CURSOR
arrows to select
position and
VALUE
arrows to select desired number (0-9).
Phase time is input as minutes and seconds within a range of 0-
49 minutes and 0-99 seconds.
5.3.2 Pre-Wash - All Units,
Alkaline Wash
1.
Press
CHANGE VALUES
. Alkaline wash phase values screen
appears:
CYCLE 10
SP=140.0 F
ALK. WASH
T=MM:SS
__Indicates flashing position.
NOTE: If treatment time is set to 0, treatment is bypassed and
question is not displayed.
To enter Wash temperature set point and phase time, press
CURSOR
arrows
to select position and
VALUE arrows
to select
desired number (0- 9). Temperature set point is input as any
number value to 1/10th of a degree, between 70 and 185ºF (22
and 85ºC). Phase time is input as minutes and seconds within a
range of 0-49 minutes and 0-99 seconds.
2.
Press
CHANGE VALUES
touch pad. Alkaline Wash Tempera-
ture Guarantee option screen appears:
GUAR. ALK. WASH TEMP?
NO
__Indicates flashing position.
Press
VALUE arrows
(either up or down) to toggle between YES
and NO answer. Selecting NO starts Wash phase time count
down at beginning of phase. Selecting YES guarantees that
Wash phase time will count down only when solution
temperature is equal to, or greater than, programmed set point.
NOTE: With Temperature Guarantee feature enabled, phase
time will not count down unless set point is reached. It is
important that set point be an attainable value. Guaranteed
temperature is shown by a G before set point (GSP instead of
SP).