6-7
Routine Maintenance
Operator Manual
122998-741
6.4 Monthly
Maintenance
6.4.1 Inspect Self-Cleaning
Filter
1.
Set
POWER-OFF/STANDBY
switch to
OFF/STANDBY
. Lock
disconnect switch in
OFF
position and close building supply
valves.
2.
Remove quick-disconnect clamp directly above pump (see Fig-
ure 6-3).
3.
Move plate to one side and remove Teflon
® 1
gasket.
4.
Remove self-cleaning filter cartridge by pulling out by handle.
5.
Rinse filter cartridge to remove any debris and inspect for dam-
age.
6.
Carefully reinsert cartridge in assembly housing and replace
gasket. Reposition plate over gasket and evenly tighten quick-
disconnect clamp.
NOTE: Gasket should last for several inspection procedures.
When necessary, order replacement gasket (P764324-398) from
STERIS
7.
Re-energize washer utilities.
1
Teflon is a registred trademark of DuPont.
WARNING – ELECTRIC SHOCK
HAZARD: Always set building
electrical-supply disconnect
switch and console master
power switch to OFF before
performing any preventive
maintenance functions within
compartments. Electrical
shock can cause serious
injury.
WARNING – BURN HAZARD:
Before performing any ser-
vice on unit, wait until cham-
ber and piping cool to room
temperature.
Figure 6-3. Self-Cleaning Filter
Assembly Housing
Self-Cleaning Filter Cartridge
Handle