Program Controller SE-707
© 2018 by STANGE Elektronik GmbH
Subject to technical modifications
47
8.1.5 Program jump
Options for program jump are displayed by operating the “Jump”
button.
Jump time
Jump to segment time
Jump set value
Jump to set value
Jump program time
Jump to program time
Jump s1
Jump to the next segment
End
Jump to program end
Controlling
Call of the control functions
Inputs which exceed the total segment time are not accepted in case of jump to segment time. The inputted time value is
balanced from segment start and the jump is executed at once. The screen automatically switches to set value curve 1, the
new segment remaining time is displayed. In case of HOLD state before the jump it is kept even afterwards; in all other cases
the program runs after jump execution (RUN).
NOTE:
Program loops are not ignored! Special attention is to be paid for programs with loops for “Jump Segment Time” and “Jump
Set Value” which should be entered with segment number. The same input of a jump may result in different targets within
the program with loops depending on actual position in program (at the beginning in reset state, right in the middle in any
segment, at the end of the program, …)!
A jump to a segment within a program loop will cause the executing of the complete loop.
8.1.6 Automatic Program Start
A start time can be entered in the RESET state of the programmer on the program curve page and on the page with the
program table via right sidebar button.
The actual loaded program in the operating memory is
automatically started exactly at the entered time (date, time) (as if
the “Start” button would be pressed).
Date and start time are displayed left next to the OPERATION
writing in inverted representation if an automatic start is
programmed.
An entered start time is deleted if:
•
the „Start off“ button is operated,
•
the program is started in other way,
•
or a program is loaded from the program memory into the
operating memory.
The entered start time can be up to 31 days in the future. Another future start time is not accepted.