General measurement functions
B:37
B
1
2
1
Determine RPM from spectrum
It is possible to determine rpm from a vibration spec-
trum when the measurement is performed without
specifying the rpm.
To activate this function, go to MENU > ‘Settings’ >
‘General Settings’ > ‘File’ tab (1) and mark ‘Possible
to determine rpm from spectrum’ (2).
Under ‘Measuring point data’ for the measuring as-
signment, the following settings are required:
• ‘Variable speed’ enabled (3)
• ‘Measure RPM simultaneously’ set to ‘No’ (4)
When you press the MEASURE/SAVE (M/S) key, you
are prompted to select how to determine rpm. Select
‘Determine rpm from spectrum’ (5).
When the measurement is completed, press F3 (‘Spec-
trum’) to go to the Spectrum window. Use the LEFT/
RIGHT arrow keys to move Cursor A (6) to the assumed
rpm (1X) in the spectrum.
In the spectrum window, press the MENU key and select
‘Set RPM from cursor’ (7). A window (8) opens where
you can verify the selected peak and the corresponding
rpm value before saving with the ENTER key.
5
4
3
6
7
8
Summary of Contents for leonova emerald
Page 1: ...User guide PORTABLE EFFICIENCY ...
Page 2: ......
Page 6: ...A 2 General instrument functions A ...
Page 36: ...B 2 General measurement functions B ...
Page 74: ...C 2 Measurement techniques always included B C ...
Page 88: ...D D 2 Shock pulse measurement ...
Page 122: ...D D 36 Shock pulse measurement ...
Page 124: ...E E 2 Vibration measurement ...
Page 150: ...E E 28 Vibration measurement ...
Page 152: ...F F 2 Rotor balancing ...
Page 170: ...F F 20 Rotor balancing ...