D
D:18
Shock pulse measurement
Measuring SPM HD
When the MEASURE/SAVE (M/S) key is pressed,
Leonova samples the transducer signal for two
seconds. A preliminary result is displayed in the
measurement window. The HDm shown is the
amplitude value of the strongest shock pulse
registered during the measuring time.
To accept the displayed result, press the ENTER
key. The measurement window shows the two shock
values (1), the status dot (2) and, if transducer line
quality test is enabled, the TLQ value (3).
Save the result by pressing the MEASURE/SAVE
(M/S) key, or measure again with SHIFT + M/S.
When ‘Variable speed’ is active, the rpm must be
measured before measuring the shock values.
3
2
1