General measurement functions
B:21
B
The graphics window shows measuring results as dots (1) against a neutral scale or, in case of evalu-
ated measuring results, a condition scale (2). Alarm limits defined in Condmaster are marked by thin
horizontal lines (3). The type of measuring result (4) is selected in the measuring result window.
Up to 100 measuring results can be downloaded with a measuring round from Condmaster. The set-
ting is made under System > Measuring system when Leonova is activated as a measuring device.
Downloading 5 to 10 measuring results is quite sufficient to see the trend when the new reading is
taken. The new result is shown before it is saved.
Graphics window
To select a measuring result dot for further examination, use the F2 function key (‘Cross’) (5). This
action displays a “crosshair” (6) which can be moved between individual measuring results using the
instrument arrow keys. The measuring result and time of measurement is displayed in the upper right
corner of the display (7).
A white corner on a function (8) indicates that more options are available when pressing the correspond-
ing function key (F1 - F4). Pressing and holding the SHIFT instrument key offers further options still.
‘Zoom’ > ‘Zoom X’ zooms the display around the center of the current view
,
while ‘Zoom to cross’
zooms in on the crosshair. ‘Zoom back’ reverts the last zoom step, while ‘Zoom back all’ returns to
the original time span.
The amplitude range can be changed by pressing SHIFT + F1. The ‘Rescale Y-axis’ function sets the
scale to the min. - max. range of the measuring results.
Comments are placed along the time line and displayed in the upper right corner (9) (replacing the
measuring result information) when put in focus by means of the crosshair. To add a new comment,
press MENU > ‘Set comment’.
Pressing MENU offers more functions, such as ‘Measuring protocol’ (10), which spaces the measuring
result dots evenly, regardless of the time intervals between measurements.
9
10
6
1
2
3
4
7
5
Comments are
displayed in the
upper right corner
8
Summary of Contents for leonova emerald
Page 1: ...User guide PORTABLE EFFICIENCY ...
Page 2: ......
Page 6: ...A 2 General instrument functions A ...
Page 36: ...B 2 General measurement functions B ...
Page 74: ...C 2 Measurement techniques always included B C ...
Page 88: ...D D 2 Shock pulse measurement ...
Page 122: ...D D 36 Shock pulse measurement ...
Page 124: ...E E 2 Vibration measurement ...
Page 150: ...E E 28 Vibration measurement ...
Page 152: ...F F 2 Rotor balancing ...
Page 170: ...F F 20 Rotor balancing ...