Fault 9: Damage has occurred to the lead screw assembly and it
cannot be removed from the press.
Cause:
Incorrect pressing of a sample.
Remedy:
Contact Specac and give serial number of press. Replace
the damaged lead screw assembly for the Atlas Power Press
from the Replacement Lead Screw Assembly Kit
P/N GS25496 – Press Service 2S-25496-1.
Fault 10: Bleeping sound and ERROR 304 message.
Cause:
Pressure Transducer not connected or faulty.
Remedy:
Contact Specac and give serial number of press. Reconnect
or replace transducer.
Fault 11: Bleeping sound and ERROR 305 message.
Cause:
On board Printed Circuit Board fuse has blown.
Remedy:
Contact Specac and give serial number of press. Obtain
replacement fuse.