Alarm Recording
An alarm recording prints the last 12 or 20 seconds (based on the
Duration
setting
found on the
Monitor Setup Printer
tab), or until the alarm ceases or is suspended,
whichever is greater.
An alarm recording for a parameter can be enabled or disabled in the Module
Configuration Manager (MCM). To enable or disable alarm recordings for those
parameters that are not available in the MCM, you will need to access the
Clinical
Setup
tab. For more information about Clinical Setup, refer to the
XPREZZON
™
and
qube
™
System Administration Manual
.
If alarm recordings are disabled for a parameter, no alarm recording produces on
either the bedside or the network printer if that parameter goes into an alarm
condition. However, the designated alarm tones still generate and the alarm
messages show.
Printing on a network printer is configuration dependent. Refer to the
XPREZZON
™
and qube
™
System Administration Manual
for details.