To turn off Full Bed Review (FBR)
1
Touch
Home
.
Alarm Watch
If your hospital protocol requires remote monitoring of all alarms, you can activate
Alarm Watch during the setup of your bedside monitor. Contact your system
administrator regarding this feature. Alarm Watch allows you to remotely watch
alarms from up to 32 networked monitors.
When a parameter from a monitor that is under
Alarm Watch
goes into alarm, a
message (for example, ALARM WATCH – BED 28 ECG CHANNEL IN ALARM) shows
on the message line of the alarm-watching monitor. The appropriate tone sounds at
the alarm-watching monitor. The corresponding parameter shows in the lowest
zone on the display of the alarm-watching monitor (unless another alarm-watched
parameter is in alarm and already occupies that zone at the watching monitor).
Note:
If the enhanced Vital Signs display feature is enabled, you can view SpO
2
,
respiration rate, noninvasive pressure, and/or invasive pressure in the ECG zone
within an Alarm Watch window. However, alarm status information for these
parameters does not show.
Warnings:
•
Alarm Watch and Remote View Alarms will generate a single tone, one time
only, at a bedside monitor.
•
Remote View and Alarm Watch are mutually exclusive features on bedside
monitors. When Remote View is active, NO Alarm Watch tones or messages
show at the bedside monitor.
•
The Module Configuration Manager settings for individual parameter
alarms could show in an Alarm Watch. Refer to the
Module Configuration
Manager System Administration Guide
for more information.
•
No more than ten remote alarms can be reported at the same time to any
monitor using the Alarm Watch feature.
•
Alarm Watch Rotation settings can delay remote alarm presentation by up
to six minutes. Biomeds verify settings and satisfactory performance during
setup.
•
The maximum number of monitors that can receive an alarm watch
notification is 16.
Caution:
When Remote View is active at a bedside monitor, no Alarm Watch alarms will
show at that bedside monitor. The Remote View alarm tone only sounds when
a remotely viewed parameter goes into alarm.