www.spacelabshealthcare.com
1 0 - 7
X P R E Z Z O N
A N D
Q U B E
O
P E R A T I O N S
M
A N U A L
C
A L C U L A T I O N S
Notes:
• Calculating VO2 and VO2I requires a CO value to complete the calculations.
The most recent CO value (less than 15 minutes old) is used for these
calculations, if it is available.
• If SvO2 is not available, ScVO2 is used if it is available. When the substitution
occurs, the following message shows: Warning ScVO2 has been substituted
for SvO2.
Output Values for Oxygenation Calculations
The following table shows the output values that are available for oxygenation
calculations.
Drug Calculator
The Drug Calculator feature lets you determine infusion rates for drugs, based on
drug amount, desired dose, and patient weight.
The following illustration shows the fields used for drug calculations.
HT
Height
cm
N/A
20 to 215
in
N/A
7.9 to 84.6
WT
Weight
kg
N/A
0.2 to 250
lb
N/A
0.441 to 551.156
Label
Name
Units
Default Value
Valid Range
Variable (Label)
Equations
Units
Normal Range
BSA
Ht
0.725
×
Wt
0.425
×
0.007184
m
2
0.03 to 3.69
Arterial Oxygen Content
(CaO
2
)
(1.34
× Hgb × SaO
2
/100) + (PaO
2
× 0.0031)
mL/dl
18 to 20
Mixed Venous Oxygen
Content (CvO
2
)
(1.34
× Hgb × SvO
2
/100) + (PvO
2
× 0.0031)
mL/dl
14 to 16
Arterial/Venous Oxygen
Difference (avDO
2
)
CaO
2
– CvO
2
mL/dl
3 to 5.5
Oxygenation Consumption
(VO
2
)
avDO
2
×
CO
×
10
mL/min
200 to 300
Oxygenation Consumption
Index (VO
2
I)
VO
2
/ BSA
mL/min/m
2
110 to 166
Partial Pressure of Alveolar
Oxygen (PaO
2
)
[(FiO
2
/ 100)
× (PB – 47)] – PaCO
2
/ 0.8
mmHg
100
Pulmonary Venous
Admixture Shunt (Qs/Qt)
100
× [(1.34 × Hgb) + (0.0031 × PaO
2
) – CaO
2
] (1.34
× Hgb) + (0.0031 × PaO
2
) – CvO
2
%
Variable, depending
on FiO
2
PaO
2
/ FiO
2
Ratio (P/F)
PaO
2
/ (FiO
2
/ 100)
mmHg or kPa
Variable, depending
on FiO
2