![Spacelabs Healthcare elance 5 Service Manual Download Page 71](http://html1.mh-extra.com/html/spacelabs-healthcare/elance-5/elance-5_service-manual_1339020071.webp)
Troubleshooting
EtCO
2
5-10
Does EtCO2
waveform appear on
display?
YES
YES
YES
End
Problem resolved
from previous
stage?
Does EtCO2 replace
RESP on display?
Problem resolved
from previous
stage?
Does EtCO2
numeric appear
correctly on display?
Replace Main board
No
No
YES
No
No
Turn on the “EtCO2 Enabled”
in Clinical menu.
No
Perform CO2 Calibration
YES
Check for occlusion.
Replace sampling line.
Re-analyze problem.
YES
Replace EtCO2
module
No
Problem resolved
from previous
stage?
Replace EtCO2
module
Connect a sampling
line
Does pump start
running?
Ensure sampling line
is compatible.
Replace sampling
line.
No
Problem resolved
from previous
stage?
YES
No
Replace EtCO2
module
EtCO2 alarms
OK?
YES
YES
No
Replace Main board
Problem resolved
from previous
stage?
YES
No
Re-analyze problem.
Summary of Contents for elance 5
Page 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Page 7: ...Introduction External Recorder 1 4...
Page 77: ...Glossary The following terms may appear in this manual 7 1...
Page 78: ...Glossary 7 2...
Page 79: ...Glossary 7 3...
Page 85: ...Appendix B Symbols B 2...
Page 86: ...Appendix B Symbols B 3...
Page 87: ...Appendix B Symbols B 4...
Page 88: ...Appendix B Symbols B 5...
Page 89: ...Appendix B Symbols B 6...
Page 90: ...Appendix B Symbols B 7...
Page 91: ...Appendix B Symbols B 8...
Page 92: ...Appendix B Symbols B 9...
Page 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...