Maintenance
Replacing the Recorder Side PCBA
1.
Disconnect battery power by moving the rear switch to off position.
2.
Remove the Front Panel.
3.
Remove the Main Board and Zbitsy Board from the rear enclosure.
4.
Remove the CO
2
PCBA if installed.
5.
Remove two screws from the middle that hold the Recorder Side PCBA to rear enclosure.
6.
Remove the recorder Side PCBA.
7.
Replace the Side Board and reassemble the monitor in reverse order.
4-30
Remove Screw
Remove Screw
Connect to Main board
Alarm LED connector