Troubleshooting
Invasive BP
5-9
Pressure numeric
appears on display?
All pressure
channels tested?
YES
YES
YES
End
Pressure waveform
appears on display?
Pressure waveform
appears on display?
Problem resolved
from previous
stage?
BP alarms OK?
Replace Main board
Replace pressure
transducer / cable
No
No
YES
No
No
Connect pressure cable to
simulator and monitor
No
Replace Main board
YES
No
Zero BP channel
Re-analyze problem.
YES
Problem resolved
from previous
stage?
No
Re-analyze problem.
YES