Maintenance
5.
Set monitor’s <NIBP Cal. Mode> to “ON “ under <Monitor Setup>/<Biomedical>/<Service>
menu.
6.
Pressurize the sphygmomanometer to 50mmHg. Observe the NIBP pressure reading on the
monitor display. Adjust the NIBP calibration potentiometer R267 until the reading is within
1mmHg of the sphygmomanometer pressure.
7.
Repeat step 6 by pressurizing the sphygmomanometer to 100mmHg and 200mmHg respectively.
8.
Release pressure and wait for NIBP reading to return to 0mmHg.
9.
Reassemble the monitor.
NIBP leakage test procedure
1.
Setup the calibration system as show below.
2.
Pressurize the sphygmomanometer to 50mmHg. Then, clamp the hose as shown above.
3.
Observe the pressure leakage during 1 minute by using a stop watch or equivalent. The leakage
rate should be within 6mmHg per minute.
4.
Repeat step 3 by pressurizing the sphygmomanometer to 100 mmHg, 150mmHg, 200mmHg and
250 mmHg respectively.
5.
If the leakage rate exceeded 6mmHg/min, then check the pneumatic system.
4-34
Clamping
position
=
Summary of Contents for elance 5
Page 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Page 7: ...Introduction External Recorder 1 4...
Page 77: ...Glossary The following terms may appear in this manual 7 1...
Page 78: ...Glossary 7 2...
Page 79: ...Glossary 7 3...
Page 85: ...Appendix B Symbols B 2...
Page 86: ...Appendix B Symbols B 3...
Page 87: ...Appendix B Symbols B 4...
Page 88: ...Appendix B Symbols B 5...
Page 89: ...Appendix B Symbols B 6...
Page 90: ...Appendix B Symbols B 7...
Page 91: ...Appendix B Symbols B 8...
Page 92: ...Appendix B Symbols B 9...
Page 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...