Maintenance
Replacing the Patient Input Panel
1.
Disconnect battery power by moving the rear switch to off position.
2.
Remove the front panel.
3.
Remove two screws that hold the patient input panel to the rear enclosure.
4.
Gently remove all the interconnection cables from the Main Board.
5.
Remove the patient input panel from the rear enclosure.
6.
Replace the patient input panel board and reassemble the monitor.
4-29
Remove Screw
Remove Screw
IBP 2
SpO
2
T1
ECG
IBP 1