Maintenance
Function Test
Setup
•
Apply Shake Test to the monitor to confirm no loose parts inside the unit.
•
Connect élance to ECG, SpO
2
and Temperature Simulators
•
Leave NIBP unconnected.
Power Up
•
Connect élance to mains power
•
Check: The yellow LED is lit.
•
Power-up the élance.
•
Check: Green LED in front panel is lit.
•
Check: Battery Charging icon is displayed (mains power)
•
Disconnect AC power from élance.
•
Check: Battery icon is displayed
S/N and SW Version Check
•
Touch <Monitor Setup> -> <Biomedical> -> <Service> -> Configuration> to bring up élance’s
Configuration Menu.
•
Check: serial number is correct
•
Check: ACQUIRE, PRESS, ECGACQ, CF displayed
•
Check: Main software version is correct
NIBP Safety and Pneumatic Checks
•
Start NIBP Measurement with hose disconnected
•
Listen for 2 NIBP pump speeds during the 1st 5 seconds of the measurement.
•
Stop NIBP Measurement.
•
Connect élance to NIBP Cuff wrapped around mandrel.
•
Start the NIBP measurement and observe the time. Squeeze the cuff every 5 seconds and then
release. Until message “NIBP measurement time exceeded” is displayed.
•
Check: NIBP measurement timeout in 135 seconds.
Main Physiological Parameters
ECG:
•
Setup monitor to display Leads III, II, V(for 5-Lead cable) or Lead II(for 3-Lead cable)
•
Verify that the selected ECG Leads are displayed correctly
•
Check parameters (HR, RR, SpO
2
, PR)
Set HR Low Alarm Limit to 60. Set HR to 40.
Verify that the ECG numeric zone turn yellow, alarm activated and message HR<60 is displayed.
Set HR High Alarm Limit to 80. Set HR to 120.
Verify that the ECG numeric zone turn yellow, alarm activated and message HR>80 is displayed.
Respiration:
•
Setup monitor to display RESP (in channel 2 only).
•
Check RESP waveform quality (clean signal),
•
Turn on the RR alarms,
•
Set RESP Lower Limit to 25.
•
Set simulator Resp Rate to 20.
•
Verify that the RESP numeric zone turns yellow, alarm activated and RR<25 message is displayed.
•
Set RESP High Alarm Limit to 35.
•
Set simulator Resp Rate to 40
•
Verify that the RESP numeric zone turns yellow, alarm activated and RR>35 message is displayed
4-5
Summary of Contents for elance 5
Page 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Page 7: ...Introduction External Recorder 1 4...
Page 77: ...Glossary The following terms may appear in this manual 7 1...
Page 78: ...Glossary 7 2...
Page 79: ...Glossary 7 3...
Page 85: ...Appendix B Symbols B 2...
Page 86: ...Appendix B Symbols B 3...
Page 87: ...Appendix B Symbols B 4...
Page 88: ...Appendix B Symbols B 5...
Page 89: ...Appendix B Symbols B 6...
Page 90: ...Appendix B Symbols B 7...
Page 91: ...Appendix B Symbols B 8...
Page 92: ...Appendix B Symbols B 9...
Page 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...