Troubleshooting
TEMP
5-8
All Temperature
channels tested?
YES
YES
End
Temperature
numeric appears
on display?
Problem resolved
from previous
stage?
Temperature
alarms OK?
Replace Main board
Replace Temperature
probe
No
YES
No
Connect temperature cable to
simulator and module
No
Replace Main board
YES
No
Problem resolved
from previous
stage?
YES
No
Re-analyze problem.