Maintenance
EtCO
2
Calibration
Required tools and parts:
1.
Oridion FilterLine sampling line
2.
5% CO
2
calibration gas cylinder.
EtCO2 calibration procedure
1. Connect an Oridion FilterLine sampling line to the élance monitor. Enable EtCO
2
monitoring in
the <Monitor Setup>/<Clinical> menu. The EtCO
2
waveform will replace RESP waveform and
display in the second waveform zone.
2. Attach the output of the 5% CO
2
calibration gas canister to the sampling line.
3. Enter the <Monitor Setup>/<Biomed>/<Service> menu and press the <Start CO
2
Cal> button.
“CO
2
calibrating” message will be displayed while calibration gas is supplied to the sampling line.
4. “CO2 calibration OK” message will be displayed if the calibration completes successfully after 1-
2 minutes. The will be an alert tone after the calibration. The EtCO
2
, FiCO
2
and RR numeric
values will return to 0.
Caution:
Dispose calibration gases in accordance with local and federal laws.
Software Update
Refer to PSN for detailed upgrade instructions.
Service Due Date Setting
Set Service Due date under
<Monitor Setup>/<Biomedical>/<Service> menu, There are three
preconfigured settings (3/6/12 months) are available.
4-35
Summary of Contents for elance 5
Page 1: ...Vital Signs Monitor 93300 05 Service Manual 070 1949 00 Rev H...
Page 7: ...Introduction External Recorder 1 4...
Page 77: ...Glossary The following terms may appear in this manual 7 1...
Page 78: ...Glossary 7 2...
Page 79: ...Glossary 7 3...
Page 85: ...Appendix B Symbols B 2...
Page 86: ...Appendix B Symbols B 3...
Page 87: ...Appendix B Symbols B 4...
Page 88: ...Appendix B Symbols B 5...
Page 89: ...Appendix B Symbols B 6...
Page 90: ...Appendix B Symbols B 7...
Page 91: ...Appendix B Symbols B 8...
Page 92: ...Appendix B Symbols B 9...
Page 94: ...Appendix B Symbols lance is a registered trademark of Spacelabs Healthcare B 11...