![Spacelabs Healthcare elance 5 Service Manual Download Page 67](http://html1.mh-extra.com/html/spacelabs-healthcare/elance-5/elance-5_service-manual_1339020067.webp)
Troubleshooting
SpO
2
5-6
YES
YES
End
SpO2 waveform
appears on display?
Problem resolved
from previous
stage?
SpO2 alarms OK?
Replace Main Board (Option S)
or
Masimo Board (Option M)
or
Nellcor Board (Option N)
Replace sensor / cable
No
YES
No
Connect SpO2 cable to
simulator and monitor
No
Replace Main board or
Masimo Board (Option M) or
Nellcor Board (Option N)
YES
Problem resolved
from previous
stage?
No
Re-analyze problem.