14
1
SIGNAL
110V/220V
~
50/60Hz
HANDLER
HV
HV
HV
HV
HV
HV
HV
MAX
5kV AC
6KV DC
MAX
5kV AC
6KV DC
MAX
5kV AC
6KV DC
MAX
5kV AC
6KV DC
MAX
5kV AC
6KV DC
MAX
5kV AC
6KV DC
MAX
5kV AC
6KV DC
MAX
5kV AC
6KV DC
HV
●
SYSTEM
Press this button to light and display the system setting interface (SYSTEM);
it is used to
configure parameters not related to test but related to the test system, such as display,
communication, etc.
3.1.14
Copy
COPY the picture of the current screen to the USB memory, which must be pre-plugged into the
front panel jack.
3.1.15
Instrument Model Label
ST9010/A Parallel Multi-Unit Hipot Tester
3.2
Rear Panel
Figure 3-2 gives a brief description of the rear panel.
12
11
10
9
8
7 6
5
2
贴标签
PAR8
PAR7
GPIB
USB
PAR6
PAR5
PAR4
PAR3
RS-23
2C
LAN
PAR2
PAR1
3
RTN/LOW
RTN/LOW
RTN/LOW
RTN/LOW
RTN/LOW
RTN/LOW
RTN/LOW
RTN/LOW
4
OUTPUT
OUTPUT
OUTPUT
OUTPUT
OUTPUT
OUTPUT
OUTPUT
OUTPUT
!
!
!
!
!
!
!
!
3.2.1
Product Name Plate
Figure 3-2 Rear panel description
Information code for product model and date of manufacture
3.2.2
Test unit output module
Parallel high-voltage output control module, 8 for ST9010 and 4 for ST9010A. Each module is
independent of each other and can test a single DUT.
Note: Do not connect different test units to the same DUT, otherwise there will be unpredictable
risks.
3.2.3
Test low terminal, test current return terminal (one per test unit)
Test low terminal of parallel high voltage test interface; all test low terminals have loop ground,
please pay attention to distinguish when conducting parallel test wiring. Wrong connection will cause
wrong judgment of the instrument and unpredictable risk.