23
the "TEST" interface.
4.2
Panel function interface and parameter description
This section introduces the functional interface and related parameters of the instrument based on the
software flow and interface correlation, to guide users to understand the function and use of the
instrument.
Instrument initial status description
1.
Turn on the instrument, it will self-test, query the existing unit and scan controller
2.
After booting, the system enters by default: TEST interface
3.
The default data is the test data when the last interface was switched before the last
shutdown.
4.
The factory default setting of the instrument is single step, AC withstand voltage, default
parameter status.
The interface switching of the instrument can be directly switched by using four function keys, they
are:
Measurement display (TEST): Start the high voltage test process.
Measurement setup (SETUP): Modify the high voltage test procedure
System interface (SYSTEM): Modify the conditions related to the instrument system
File processing (FILE): Save and recall the set test file from the instrument storage area or USB
The basic function description of the button:
Note:
The instrument will
save the data modification only
when the
interface is switched.
After modifying the data in the setup interface and system interface, you must switch the interface.
Please be careful not to
shut down directly
after modifying the data to avoid
data loss
.
TEST (test key):
Let the instrument enter the test wait state, ready to start
the high voltage test. (Power on defaults to this interface)
SETUP
(
setup key
):
This is the interface used to modify the current test file,
test project, test parameters. The modification of the test
file is completed in this interface.
SYSTEM (system key):
Test safety and instrument operating mode related settings,
it has little to do with test files. Press multiple times to
switch the sub-interface.
FILE (file key):
Test file save and call interface, it is related to data storage.
Press multiple times to switch the sub-interface.
Note:
Call may have a version mismatch problem.
(direction key):
In this interface, press the cursor up, down, left, and right.
Knob:
In this interface, move the cursor quickly in order.
F1~F6 (soft key):
Work with the right area contents of the soft key function
displayed to achieve quick modification.
0~9 (
numeric keypad):
Modify the data at the cursor position, please refer to the
information prompt area
at the bottom of the screen to
determine the data input range.