51
Commands for DC Setup Function
FUNC
:
SOURce
:
STEP <sn>
:
DC
:
WTIM
To set /inquiry about the wait time for DCW test.
(
Syntax refer to
AC
:
TTIM)
FUNC
:
SOURce
:
STEP <sn>
:
DC
:
RAMP
To set/query the ramp state of DCW.
--
Syntax:
Command message:
FUNC
:
SOUR
:
STEP <sn>
:
DC
:
RAMP
:
<ON/OFF> or <1/0>
Query message:
--Example:
FUNC
:
SOUR
:
STEP <sn>
:
DC
:
RAMP
:
?
Set the status of RAMP of DCW in STEP 1 as
:
ON
Command message: FUNC
:
SOUR
:
STEP 1
:
DC
:
RAMP ON
--Return Message
Query message: FUNC
:
SOUR
:
STEP 1
:
DC
:
RAMP?
Return to RAMP status of DCW in STEP 1, such as ON
Commands for IR Setup Function
FUNC
:
SOURce
:
STEP <sn>
:
IR
:
UPPC
To set /inquiry about the UPPER resistance for
IR test.
--
Syntax:
Command message
:
FUNC
:
SOUR
:
STEP <sn>
:
IR
:
UPPC <resistance >
Query message
:
FUNC
:
SOUR
:
STEP <sn>
:
IR
:
UPPC?
--
Data< resistance>:
Data format: float
Data range: 0, 0.1~10000MΩ
(0 is OFF)
Data accuracy: 0.1 MΩ
Data unit: MΩ
--
Example:
Set the upper resistance for IR test in STEP1 to 1 MΩ.
Command message: FUNC
:
SOUR
:
STEP 1
:
IR
:
UPPC 1
--Return Message:
Query message: FUNC
:
SOUR
:
STEP 1
:
IR
:
UPPC
Return to upper limit of IR in STEP 1, such as 1
FUNC
:
SOURce
:
STEP <sn>
:
IR
:
LOWC
To set /inquiry about the LOWER resistance for IR
test.
(
Syntax refer to
IR
:
UPPC)
FUNC
:
SOURce
:
STEP <sn>
:
IR
:
RANG
To set /inquiry about the RANG for IR test.
--Syntax:
Command message: FUNC
:
SOUR
:
STEP <sn>
:
IR
:
RANG <rang>
Query message:
FUNC
:
SOUR
:
STEP <sn>
:
IR
:
RANG?